Operation Manual

For

NDJ-1 Rotary Viscometer

MADE  IN  CHINA

Thank you for using NDJ-1 Rotary Viscometer. NDJ-1 is easy to use, in order to install and operate accuray, and make full use of product features, we suggest you to read the manual before using.

· The company possesses the final interpretation power.

· The company reserves the power of amending technical rules without prior notice.

· The company reserves the power to amend the manual without further notice.

· May not be translated into other languages, reproduction prohibited without written permission. 

               

               Cataloge

1General……………………………………………………

2Main technical data…………………………………………

3Principle………………………………………………

4Installation……………………………………………………

5Operation procedures………………………

6Precaution………………………………………………

Packing List ……………………………………………………

1. General

NDJ-1 rotary viscometer is a new instrument used for determining the liquid viscose capacity and the absolute viscosity. NDJ-1 has been widely used to determine and measure the liquid viscosity in many applications such as grease, painting, plastic, pharmacy and adhesives. It is a precision instrument to monitor and control the stable quality of products in the manufacturing.

2. Main technical data

1) Measurement range: 10-100.000 mPa·s 

2) Rotor specification: 0, 1, 2, 3, 4, 5, five kind rotor (0# rotor could measure the viscometer from low viscometer to 0.1 mPa·s)

3) Measurement error: ±5% (Newton liquid)

4) Power supply: Voltage--220±22V, Frequency50 Hz±0.5Hz

5) Dimensions: 300 mm ×300 mm × 450 mm

6) Net weight: 1.5 Kg (the base not included)

3. Principle

1) Figure 1Synchronous motor rotates with stable rate, connect scale disk, through hairspring and rotating shaft, drives rotors to rotate. The rotors will subject to a torque moment proportional to liquid viscosity because of the liquid viscose hysteresis. The torque moment will be measured by the sensors and processed into the viscosity and shown on the display.

2) Rate can be changed by using gear system and clutch, operated by rotary knob, four optional velocity.

3) According to different equipments, different rotors (Five rotors: NO.0-NO.4) will be attached, it can be select with the velocity according to the viscosity of the liquid.

4) The fixed control gear of pointer are used for accurate reading. When the rate is so fast (30 rpm, 60 rpm) that you can not get the reading while it rotated, press pointer joystickfix the pointer in order to get the reading.

5) The rotor bracket is used for the protection of rotors and the stability of the measurement, a more reliable measuring result can be made by using the protection bracket. 

6) It can be used as a portable instrument, assigned fixed bracket and elevating system. Generally, it should be fixed when determine small amount or the temperature is fixed in the laboratory.

7) The instrument has gears and damping elevate mechanism. It is stable for instrument lifting.

8) The whole set of instrument has good quakeproof packing to ensure the precision of instrument.

Installation

1. Take out the bracket, package, stanchion and so on from the packaging.

2. Screw the column to the threaded hole behind the bracket, and put the tooth profile surface in the column facing front the bracket. Screw the hex nut tight by wrench to prevent the column running.

3. Rotational move the lifting chuck knob to check the flexibility and self-looking of lifting chuck. Found the lifting chuck too loosen and too tight phenomenon, can be used hex nut wrench to adjust the tight screw. So that it can lift from top to bottom. And it is suitable for tight to prevent the situation of automatically falling after installing the viscometer.

4. Take out viscometer form package and mount it on the lifting collettight it by handle setscrew to avoid loosen(as level as possible), link the power supply, take away the rubber band on the pointer joystick, screw the nut of protecting-cap loosen under the instrument, take away the protecting -cap.

5. Adjust the adjustment screw and put the rotor into the liquid to be measured till the level mark on the rotor reach the liquid surface.

Operation procedures

1. Prepare the liquid to be measured and put it into a glass beaker or a right angle container with the diameter not smaller than 70 mm, height not smaller than 130mm, take care of the liquid temperature.

2. Mount the protection bracket on instrument, turning right for mounting, turning left for removing.

3. Screw the selected rotor into connecting rod, turning right for mounting, turning left for removing , Adjust the lifting screw and put the rotor into the liquid to be measured till the level mark on the rotor reach the liquid surface. Connect power supply, turn on the equipment, screw the speed knob, select velocity, relax pointer joystick, reading data can be displayed (about 20~30 seconds) when the pointer becomes stable. Press the pointer joystick after pointer becoming stable, cut power supply, then get the reading data. If the pointer isnt in reading window after motor stop, you can press pointer joystick unceasingly, open or close the motor time after time, after practice, you can operate it familiarly.

4. When the reading data is too high or too low, the rotor or the rate can be changed, make sure the reading is between 20 and 80.

5. Choice of range, coefficient, rotor and rate:

5.1 Estimate approximay the viscose range, then select the rotor and the velocity based on the range table given below:

For example, the viscosity of liquid is about 3000 mPa·s the settings should be as following:  

rotor 2 # with velocity of 6 rpm,

rotor 3 # with velocity 0f 30 rpm

5.2 If the viscosity of liquid can not be estimated, it should take a high value for real measurement, the rotor selection should be made from smaller to larger with their number from higher to lower. In general, high viscosity should use smaller rotor with slower velocity, lower viscosity use lager rotor with faster velocity.

5.3 Coefficient: the reading must be multiplied the specific coefficient in the coefficient table in order to get the absolute viscosity.

   That is: 

            absolute viscosity

            coefficient

            reading ( deflection angle)

   5.4 Amendment of frequency error: when the power frequency is inaccurate, you can amendment it according to the formula below:

      Actual viscosity= instruction viscosity*nominal frequency/ actual frequency

Velocity

Range

Rotor

60

30

12

6

1

100

200

500

1000

2

500

1000

2500

5000

3

2000

4000

10000

20000

4

10000

20000

50000

100000

   5.5 

5.6 Coefficient table: 

        Rpm

Rotor

60

30

12

6

1

1

2

5

10

2

5

10

25

50

3

20

40

100

200

4

100

200

500

1000

 

Precaution

1. This instrument is limited to the room temperatures

2. This instrument should be used under the designed voltage and frequency and their allowable error ranges, or incorrect results could be resulted.

3. Using the bracket fix the instrument in order to determine accuray. Handheld operation: keep the instrument stable and level.

4. Care should be taken for mounting or removing rotors, slightly and lift the connecting screw bolt to avoid a transverse force acting on rotor to cause it bending.

5. The instrument mounted with rotor should not be revolved.

6. Don’t run the motor when the pointer joystick is not pressed.

7. Keep cleaned on screws and connecting points between rotor and connecting bolt rod, or a unstable rotation could be caused in the real measurement.

8. Use hand to hold the instrument when lifting up or moving down to avoid dropping it.

9. After completing measurement each time, the rotor should be fully cleaned (rotor should be removed from instrument for cleaning), then place it on the protection bracket.

10. Unauthorized removing or replacing the instrument part, and applying lubricates are not allowable.

11. When moving or shipping instrument, put on the yellow caver cap and lift the connecting bolt rod and screw the bolt on the cap tightly.

12. Suspension, emulsion or polymer and other high viscosity liquids are non-Newton liquids, their viscosity will change with shear velocity and time, it is normal for their measured inconsistent results under the selected rotor and velocity and time, and it is not resulted from the instrument problems (in general, the rotor and velocity and time should be specified for non-Newton liquids).  

13. The cautions should be taken for followings to obtain a good measuring reault:

(1) Accuray control the temperature of liquid to be measured;

(2) Make temperature of rotor and protection bracket same as liquid;

(3) Ensure the liquid homogeneity and without bubbles;

(4) Put the rotor located on the center of the liquid container in real measurement;

(5) Remove bubbles adhered on the rotor when put it into liquid;

(6) When change the rotor or rate, keep between 30~90 as possible.

(7) Use rotor protection bracket for measurement;

(8) Ensure rotor and connecting screw rod cleaned;

(9) Amend according to the amending formula when power frequency is not exact.

(10) When pointer is stable, reading with the same angle of view.

(11) Adjust level bubble exactly, move the instrument slowly, keep the instrument stable.

(12) Strictly follow the operation instruction for measurement.

Packing List

1) NDJ-1 Digital Rotary Viscometer           1 Qt.

2) Rotors, 1#, 2#, 3# and 4#                   1 Qt of each

3) Protection bracket                       1 Qt.

4) Bracket                                1 Qt.

5) Adjusting screw                         3 Qt.

6) Operation manual                        1 Qt.

7) License                                1 Qt.

8) Warranty card                           1 Qt.

上海平軒科學儀器有限公司

上海平軒科學儀器有限公司

工廠地址:上海市嘉定區南翔德園路1259弄18號

©2018 版權所有:上海平軒科學儀器有限公司  備案號:滬ICP備10019381號-1  總訪問量:1568830  站點地圖  技術支持:化工儀器網  管理登陸

主站蜘蛛池模板: PO膜_灌浆膜及地膜供应厂家 - 青州市鲁谊塑料厂 | 天津热油泵_管道泵_天津高温热油泵-天津市金丰泰机械泵业有限公司【官方网站】 | 房车价格_依维柯/大通/东风御风/福特全顺/江铃图片_云梯搬家车厂家-程力专用汽车股份有限公司 | 超声波成孔成槽质量检测仪-压浆机-桥梁预应力智能张拉设备-上海硕冠检测设备有限公司 | led全彩屏-室内|学校|展厅|p3|户外|会议室|圆柱|p2.5LED显示屏-LED显示屏价格-LED互动地砖屏_蕙宇屏科技 | 粤丰硕水性环氧地坪漆-防静电自流平厂家-环保地坪涂料代理 | 数码听觉统合训练系统-儿童感觉-早期言语评估与训练系统-北京鑫泰盛世科技发展有限公司 | 北京发电车出租-发电机租赁公司-柴油发电机厂家 - 北京明旺盛安机电设备有限公司 | 东莞压铸厂_精密压铸_锌合金压铸_铝合金压铸_压铸件加工_东莞祥宇金属制品 | 广东燎了网络科技有限公司官网-网站建设-珠海网络推广-高端营销型外贸网站建设-珠海专业h5建站公司「了了网」 | 外贮压-柜式-悬挂式-七氟丙烷-灭火器-灭火系统-药剂-价格-厂家-IG541-混合气体-贮压-非贮压-超细干粉-自动-灭火装置-气体灭火设备-探火管灭火厂家-东莞汇建消防科技有限公司 | 凝胶成像仪,化学发光凝胶成像系统,凝胶成像分析系统-上海培清科技有限公司 | 美侍宠物-专注宠物狗及宠物猫训练|喂养|医疗|繁育|品种|价格 | 回转炉,外热式回转窑,回转窑炉-淄博圣元窑炉工程有限公司 | 智能家居全屋智能系统多少钱一套-小米全套价格、装修方案 | 山东聚盛新型材料有限公司-纳米防腐隔热彩铝板和纳米防腐隔热板以及钛锡板、PVDF氟膜板供应商 | sus630/303cu不锈钢棒,440C/430F/17-4ph不锈钢研磨棒-江苏德镍金属科技有限公司 | 分子蒸馏设备(短程分子蒸馏装置)_上海达丰仪器 | 超高频感应加热设备_高频感应电源厂家_CCD视觉检测设备_振动盘视觉检测设备_深圳雨滴科技-深圳市雨滴科技有限公司 | 齿轮减速机电机一体机_齿轮减速箱加电机一体化-德国BOSERL蜗轮蜗杆减速机电机生产厂家 | 药品冷藏箱厂家_低温冰箱_洁净工作台-济南欧莱博电子商务有限公司官网 | 自恢复保险丝_贴片保险丝_力特保险丝_Littelfuse_可恢复保险丝供应商-秦晋电子 | POM塑料_PBT材料「进口」聚甲醛POM杜邦原料、加纤PBT塑料报价格找利隆塑料 | 衬塑设备,衬四氟设备,衬氟设备-淄博鲲鹏防腐设备有限公司 | 百方网-百方电气网,电工电气行业专业的B2B电子商务平台 | 多功能三相相位伏安表-变压器短路阻抗测试仪-上海妙定电气 | 煤矿人员精确定位系统_矿用无线通信系统_煤矿广播系统 | 流变仪-热分析联用仪-热膨胀仪厂家-耐驰科学仪器商贸 | 环氧乙烷灭菌器_压力蒸汽灭菌器_低温等离子过氧化氢灭菌器 _低温蒸汽甲醛灭菌器_清洗工作站_医用干燥柜_灭菌耗材-环氧乙烷灭菌器_脉动真空压力蒸汽灭菌器_低温等离子灭菌设备_河南省三强医疗器械有限责任公司 | 沈阳楼承板_彩钢板_压型钢板厂家-辽宁中盛绿建钢品股份有限公司 轴承振动测量仪电箱-轴承测振动仪器-测试仪厂家-杭州居易电气 | 耐磨陶瓷管道_除渣器厂家-淄博浩瀚陶瓷科技有限公司 | 食药成分检测_调料配方还原_洗涤剂化学成分分析_饲料_百检信息科技有限公司 | 煤粉取样器-射油器-便携式等速飞灰取样器-连灵动| 生态板-实木生态板-生态板厂家-源木原作生态板品牌-深圳市方舟木业有限公司 | 中医治疗皮肤病_潍坊银康医院「山东」重症皮肤病救治平台 | 铝镁锰板厂家_进口钛锌板_铝镁锰波浪板_铝镁锰墙面板_铝镁锰屋面-杭州军晟金属建筑材料 | 视频直播 -摄影摄像-视频拍摄-直播分发 | 阀门智能定位器_电液动执行器_气动执行机构-赫尔法流体技术(北京)有限公司 | 工业机械三维动画制作 环保设备原理三维演示动画 自动化装配产线三维动画制作公司-南京燃动数字 聚合氯化铝_喷雾聚氯化铝_聚合氯化铝铁厂家_郑州亿升化工有限公司 | 粉末包装机-给袋式包装机-全自动包装机-颗粒-液体-食品-酱腌菜包装机生产线【润立机械】 | ge超声波测厚仪-电动涂膜机-电动划格仪-上海洪富 |